|
Standard Operating Procedures
Power on the system beginning with the D-7000 interface.
Open HSM software and click the “System Status” icon on the left
toolbar. Select “Initialize.” A connection with the instrument is set
up and the program is downloaded. This may take approximately 2 minutes
to complete. Make sure the hardware detected by the computer is
correct.
Purge each solvent line for 2-3 minutes by following this procedure:
- Hit the “Manual Set” button on the L-7100 pump
- Open the white purge valve to the right of the pumps
- Hit the “Pump On” button
- Hit the “Purge” button to start and again to stop the manual purge
- Hit the “Pump On” button to turn the pumps off
- Close the purge valve to allow solvent flow to the column
Setting up a Method Table
***Work left-to-right on the icon list at the top of the screen***
First row icons set acquisition parameters. Second row icons are
data reprocessing functions.
Method Configuration Icon
Verify the following:
|
Interface Module |
D-7000 I/F |
|
Gradient Mode |
LOW |
|
Channel 1 Detector |
L-7455 for DAD |
|
Channel 2 Detector |
L-7485 for FL |
|
Column |
C18 |
|
Column Oven |
NONE |
|
External Instrument Software |
NONE |
|
Autosampler |
L-7200 |
|
Pump A |
L-7100 |
|
Pump B |
NONE |
|
Number of Solvents |
One to four |
Method Information Icon
This window allows you to provide a description of the method, the
name of the developer, and the solvents used. The method name is
displayed as a read-only field.
Pump Setup Icon
Setup the solvent profile for your run here. Do not change the
pressure limit values.
Autosampler Setup Icon
Use to specify syringe parameters, injection method, and needle
wash. Leave the default values for needle and injection port washing.
The L-7200 autosampler permits three injection methods:
- Cut Injection Method – the amount of sample
withdrawn by the needle is slightly larger than the amount that will
be transferred from the injection valve to the column. The middle
part of the sample is injected; the leading and trailing ends are not
injected. This method provides good reproducibility, however some of
the sample is wasted.
- All Volume Injection Method – the entire sample
withdrawn by the needle is transferred to into the sample loop and
into the column. This method is suitable for analysis of minute
samples since no part of the sample is washed.
- Full Loop Injection Method – the sample loop is
filled with sample for measurement. This method provides the best
repeatability of sample injection volume in analysis.
Syringe capacity: 0.5 ml (standard syringe)
Sample injection volume: 0.5 ml to 400 ml (standard loop: 100 ml)
pH range: 1 to 13
Channel 1 Detector Setup Icon
For the diode array detector (L-7455), you can specify the
following:
- spectra properties such as slit width, spectral bandwidth, and
spectral interval
- wavelength range and monitoring wavelength
- start and stop times
- autozero the absorbance before injection
- performance check before series start
The default spectral bandwidth is Auto and the default spectral
interval is 1600 msec.
Channel 2 Detector Setup Icon
For the fluorescence detector (L-7485), you can specify the
following:
- time period over which data are averaged – the default is 2.0 – a
longer response time gives better noise filtering while a shorter
response time gives better sensitivity
- PMT voltage (Low, Medium, or High)
- emission bandwidth – select Standard (15nm slit) or Wide (30nm
slit)
- if you want to set up the Sampling Period Table
Event Signal Setup Icon
Used if an external output is connected to the D-7000 Interface to
receive signals. Do not use.
Time Summary Icon
This window displays the solvent properties and gradient over time
specified in the Pump Setup section.
NOTE: The second row of Method icons are related to data
reprocessing and are briefly described below.
Calculation Method Icon
Options for data quantitation and identification can be specified
here.
Component Table Icon
Used to set up, review, or modify the component table for data
processing.
Integration Table Icon
Specifies how the baseline for the chromatogram is determined and
how the peaks are detected.
DAD Data Processing Icon
Specifies DAD detector processing properties such as peak purity
checks, setting the type of chromatogram to display, and other data
options.
Chrom Display Format Icon
Specifies how the chromatogram will be displayed on the Data Acquisition
and Data Processing screens as well as in a report.
DAD Display Format Icon
Use this command to display and modify the DAD display format.
Confidence Report Icon
Used to set report parameters for System Sustainability Tests, Module
Performance Tests, and Data Diagnosis.
Report Format Icon
Used to select report options, including summaries, statistics, report
layout and destination, as well as reported peaks.
Update Method Icon
Use this command to update all parameters of the current method to all
other open windows. This command is particularly useful if you are
modifying the data processing or displaying parameters in the current
method and want to update the changes to the open Data Display window.
REPORTS
Two types of reports:
Original report – raw data – this report never changes
Modified report – any changes made to an original report are saved as a
modified report. You can create any number of modified reports,
however, the report is overwritten each time you save.
Starting Data Acquisition
Open a Sample Table for data acquisition and click the eyeball icon
on the left tool bar. The “Idle Monitor” window will appear. (Make
sure the system has been initialized with the proper components!)
Manually turn on the pumps.
Click on the “Start Series” button to begin data acquisition.
***Make sure to set system to SLEEP MODE if your run will continue
overnight*** |